📊 HPLC Calculations
Calculate all five key HPLC system suitability parameters — retention factor, resolution, theoretical plates, tailing factor and assay.
System suitability
All 5 parameters required for ICH and USP system suitability testing.
Pass/Fail guidance
Each result shows whether it meets standard acceptance criteria.
Always free
No sign up, no limits. Use for every HPLC run.
HPLC system suitability parameters: interpretation guide
USP General Chapter <621> defines the system suitability parameters required before and during validated HPLC runs. Theoretical plates N ≥ 2000 for most analytical applications. Tailing factor T ≤ 2.0 (USP). Resolution Rs ≥ 1.5 for baseline separation of adjacent peaks. These values must be measured from the same injection used for the sample, not from a separate calibration run. Report all SST parameters in your analytical method validation report alongside the acceptance criteria from your method SOP.